IMR 265, High Pressure Hot Cathode, DN 25 ISO-KF
Flange size: DN 25 ISO-KF
Measurement range from 2 · 10-6 to 1000 hPa
Two gauge heads (Pirani and hot cathode) in a single case
Highly accurate
Excellent reproducibility
Automatic cathode protection
A single flange from atmosphere to UHV
Corrosion-resistant
Maximum pressure refers to inert gases and temperatures of less than 55 °C
Technical Data IMR 265, High Pressure Hot Cathode, DN 25
ISO-KF
Accuracy: % of measurement 10-5- 1 hPa: ± 15 %
Bakeout temperature 150 °C|302 °F|423 K
Electron collector Stainless steel
Filament Iridium yttriated
Filament holder Molybdenum, Platinum
Flange Stainless steel
Flange (in) DN 25 ISO-KF
Insulator Glass
Ion collector Stainless steel
Measurement range max. 1,000 hPa|750 Torr|1,000 mbar
Measurement range min. 2 · 10-6hPa|1.5 · 10-6Torr|2 · 10-6mbar
Output signal: Minimum load 10 k
Output signal: Pressure range Ioni 1.5 - 7.5 V
Output signal: Pressure range Pirani 8.5 - 9.75 V
Output signal: Sensor error 0.5 V
Pirani measurement element Copper, Tungsten
Pressure max. 5,000 hPa|3,750 Torr|5,000 mbar
Repeatability: 10-1- 102hPa 30 % reading
Repeatability: 10-5- 10-1hPa 2 % reading
Sensor cable length 100 m
Supply: Power consumption max. 16 W
Supply: Voltage 20-30 V DC
Temperature: Operating 0-50 °C|32-122 °F|273-323 K
Temperature: Storage -20-70 °C|-4-158 °F|253-343 K
Volume 20 cm3
Weight 285 g