Brand: Agilent /HP
Frequency range: 300 KHZ ~ 3 g / 6 GHZ
Mixer integrated 1 hz resolution composite signal source
In pieces with external disk drive storage/call directly
Mixer time-domain analysis
Pieces of test sequence function is used to implement complicated test process
Pieces of 100 db dynamic range
Mixer can measure the group delay and with linear phase deviation
0.001 dB, 0.001, 0.010 ns markers of resolution
Mixer increase built-in precision
Harmonic mixer scanning measurement
describe
HP8753C/D network analyzer for laboratory and production test field provides superior RF network measurement equipment. As the combination of a set of test device, it for the analysis of 300 KHZ ~ 6 GHZ active and passive network, devices and components of linear feature provides a good solution. Network analyzer has two independent display channel, you can simultaneously measure the reflection and transmission characteristics of the device under test, measurement and separation on the surface of the screen to overlap or observation results. Ease of use of preprogrammed buttons enable you to quickly measure amplitude, phase, or of the device under test group delay characteristics.
Test the timing function allows a type to quickly and repeatedly perform complex tasks. Under the timing works, from the panel to measure it only once and analyzer can store type, so that no additional programming. Can also be used to test sequence by HP - IB port to control other devices. Other efficiency measures including plotter/printer buffer, the limit test and tag, arbitrary frequency tracking function. Piecewise calibration and interpolation error correction can improve school HP8753C analyzer has a certain section of the frequency range of vector precision.
Integrated synthesis source to provide greater than 100 mw of power output, 1 hz frequency resolution and linear, logarithmic, lists, power and continuous wave scan type. Three tuning receiver on 300 KHZ ~ 3 GHZ (option 006 extended to 6 GHZ) in more than 100 db dynamic range of an independent power measurement or ratio measurement at the same time. When HP85047A HP8753C connection test device, can from 300 KHZ ~ 3 GHZ or from 3 MHZ ~ 6 GHZ (test device with frequency multiplier) research reflection and transmission characteristics of the device under test.
Nonlinear device testing
HP8753C has the measuring of the capability of nonlinear device characteristics, when increasing the function of harmonic measurement (option 002), can be directly or relative to the basic carrier (dBc) shows the amplifier scan, twice and three times harmonic level. When a scan as basic signal frequency is lower than the 16 MHZ, can quickly and convenient measurement of 40 DBC harmonic amplifier, using the same test equipment is used to measure the gain. Power meter calibration relatively absolute input or output level sensitive devices provide the absolute power of the image. HP8753C automatically on the HP436A, HP437B, or HP438A power meter, in any place in the test system of power is transferred to a power meter accuracy. Analyzer also complete mixer series harmonic frequency conversion loss measurements. Can also fixed the IF and scan the IF measurement.
The time domain analysis
HP 8753 c (010) with option to display the function of the network time domain response, the analyzer to calculate frequency response of Fourier inverse transformation, according to the network of the reflection or transmission coefficient with time.
HP 8753 c provides two time domain model, low-pass patterns provide the traditional time domain reflectometry (TDR) measurement, and gives the step response and the shock response of the network. This model is given in the impedance of the discrete state (R, L, C) type of information. Bandpass time domain model, only pulse excitation, there is no limit to the frequency, and the time domain response of frequency selective equipment, such as a SAW filter or antenna. Gating function can be used to selectively isolate a response signal, and observe a don't do their own part of a single part of the frequency domain response of the interference.
HP 8753 c general technical indicators
Signal source frequency characteristic
Scope: 300 KHZ ~ 3 GHZ
Resolution: 1 hz
Accuracy (25 ℃) : plus or minus 10 PPM
The output characteristics of power range: 5 ~ + 20 DBM
Level precision (50 MHZ, + 10 DBM) : + / - 0.5 dB
Linear: level - 5 ~ 0 DBM: + / - 0.5 dB; 0 ~ 15 DBM: + / - 0.2 dB
+ 15 ~ + 20 DBM: + / - 0.5 dB
50 Ω impedance:
Harmonic: 25 or less DBC (20 DBM output level)
- 50 or less DBC (0 DBM output level)
Not related harmonic mixer harmonic stray signals:
Or less - 32 DBC (20 DBM output level)
- 55 or less DBC (0 DBM output level)
Other spurious signals: f < 135 MHZ: 60 DBC
F 135 MHZ or higher: 60 dBc + 20 * log (f / 135 MHZ) dBc
Phase noise (1 hz bandwidth, 10 KHZ offset) : f < 135 MHZ: - 135 DBC
F 135 MHZ or higher: 135 dBc + 20 * log (f / 135 MHZ) dBc
receiver
Frequency range: 300 KHZ ~ 6 GHZ
Type: type A, B, 100 db dynamic range < 3 GHZ, 95 db dynamic range 3 ~ 6 GHZ
Sensitivity (noise level) :
3 KHZ bandwidth: - 90 DBM < 3 GHZ. - 85 DBM 3 ~ 6 GHZ
Hz bandwidth: 10-100 DBM < 3 GHZ, - 95 DBM 3 ~ 6 GHZ
Maximum input level: 0 DBM
50 Ω impedance:
Input cross interference:
300 KHZ ~ 1 GHZ: - 100 db
1 GHZ ~ 3 GHZ: - 90 db
3 GHz ~ 4.5 GHz: - 85 db
4.5 GHz ~ 6 GHz: - 75 db
Dynamic accuracy: + / - 0.05 dB, plus or minus 0.3 o
Group delay characteristics
Scope: 1 / (2 * minimum aperture)
Aperture: optional
20% of the largest: frequency interval
Minimum: frequency interval)/(point 1)
Resolution: 27.8 / aperture (Hz)
The typical value of 0.01