Definitions
General
Product data applies under the following conditions:
• Three hours storage at ambient temperature followed by 60 minutes warm-up operation
• Specified environmental conditions met
• Recommended calibration interval adhered to
• All internal automatic adjustments performed, if applicable
• Unless stated otherwise, specifications apply to test ports and a nominal source power of –10 dBm
Specifications with limits
Represent warranted product performance by means of a range of values for the specified parameter. These specifications are
marked with limiting symbols such as , ≥, ±, or descriptions such as maximum, limit of, minimum. Compliance is ensured by
testing or is derived from the design. Test limits are narrowed by guard bands to take into account measurement uncertainties, drift
and aging, if applicable.
Specifications without limits
Represent warranted product performance for the specified parameter. These specifications are not specially marked and represent
values with no or negligible deviations from the given value (e.g. dimensions or resolution of a setting parameter). Compliance is
ensured by design.
Typical data (typ.)
Characterize product performance by means of a representative value for the given parameter (e.g. nominal impedance). In contrast to
typical data, a statistical evaluation does not take place and the parameter is not tested during production.
Measured values (meas.)
Characterize expected product performance by means of measurement results gained from individual samples.
Uncertainties
Represent limits of measurement uncertainty for a given measurand. Uncertainty is defined with a coverage factor of 2 and has been
calculated in line with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into account
environmental conditions, aging, wear and tear.
Device settings and GUI parameters are indicated as follows: “parameter: value”.
Typical data as well as nominal and measured values are not warranted by Rohde & Schwarz.
Version 03.02, October 2014
4 Rohde & Schwarz R&S
ZNC Vector Network Analyzer
Measurement range
Impedance 50 Ω
Test port connector N female
Number of test ports 2
Frequency range R&S
ZNC3 9 kHz to 3 GHz
Static frequency accuracy 8 ppm
Frequency resolution 1 Hz
Number of measurement points per trace 2 to 5001
Measurement bandwidth 1/1.5/2/3/5/7 steps 1 Hz to 300 kHz
Dynamic range 1 9 kHz to 100 kHz > 90 dB, typ. 110 dB
100 kHz to 1 MHz > 115 dB, typ. 120 dB
1 MHz to 3 GHz > 120 dB, typ. 130 dB