Safety standards
- IEC60601-1:1988 IS09919:2005(E) IEC60601-2 - Type of protection; lnternally powered equipment
Degree of protection; Type BF-1o,pplied part
Measurement atlow perfusion
The values of Sp02 and pulse rate can be displayed properly
when pulse saturation is at 0.3%
Resistance to surroundinglight
The difference between the value measured in the condition of indoor natural light and that of darkroom isless than1%
Resistance to interference of man-made light
The dilference between the value measured in the condition of man-madelight and that of darkroom isless than1%
Automatic shutdown function
The device will power olfautomatically in8seconds when there is no finger in the device.
Display
M70C Display Format: OLED
Resolution: 128*64
Parameter: SP02 Pulse Rate, Pulse intensity BarGraph,,battery capacity indication, Pleth Waveform
Display Mode;Fourdirections, short press power button can change direction ofthe display
Bl_T Digital Sp02 (Pulse Oximetry)
Range:0~100%
Resolution: 1%
Accuracy: 2% (70%~100C%l1)
Unspecified(<70%
Pulse Rate
Measurementrange: 25bpm~250bpm
Resolution: lbpm - Accuracy: 1%or1 BPM(whichever is bigger)
Power
Powersupply requirment: 2AAAAlkaline batteries - Operating current: <40mA - Life:Approximate24hours continuous working time
Environmental
Temperature:
Operation: 0~+40
Storage: _20~+55
Humidity:
Operation: 0%-85%,noncondensing
Storage: 0%-93%,nonoondensing
Atmospheric Pressure;
Operation: 860hPa~1060hPa
Storage: 500hPa~1060hPa
Standard Accessory
- 2AAAAlkaline batteries
- operating manual
- Hang string